Model 6487 Firmware Upgrade Notice for KE6485 IVI/VISA Instrument Driver Support and ExceLINX (ELNX-852C02),A Microsoft Excel Add-In Supporting Keithley 6485, 6487 and 6514 Meters. 4/02/2003 Introduction The Model 6487 requires firmware revision A02 to work properly with the KE6485 IVI/VISA Instrument Driver and ExceLINX (ELNX-852C02), If your instrument's firmware revision is lower than A02, a firmware upgrade is required. This upgrade can also be used for those who wish to have the latest firmware enhancements and upgrades installed on their instrument. Step 1. Determine the firmware revision of your Model 6487 The firmware revision is briefly displayed during the power-on sequence of the instrument. When the instrument's power is turned on, all display annunciators will be illuminated for a short period. Then the firmware revision is displayed briefly as follows: 6487: yyy where: yyy is the firmware revision (e.g. A01) of your Model 6487. To determine the firmware revision with remote control, use the *IDN? query command. After sending the *IDN? command and addressing the instrument to talk, the following response is sent to the computer: KEITHLEY INSTRUMENTS INC., MODEL 6487, xxxxxxx, yyy/zzz Where: xxxxxxx is the serial number, yyy is the firmware revision, and zzz is the display firmware revision. Step 2. Determine next action Based on the firmware revision of your Model 6487, choose the appropriate course of action: If the firmware revision is A02 or higher, an upgrade is not required to use the KE6485 IVI/VISA Instrument Driver and ExceLINX (ELNX-852C02) with your instrument. If the firmware revision is A01, an upgrade is required. Proceed to step 3 to perform the upgrade. Step 3. Upgrade firmware To perform the firmware upgrade, a personal computer connected to your 6487 is required. The PC must be at least Pentium class running Windows 98 or later. The firmware can be installed via the IEEE-488 bus or the RS-232 interface: IEEE-488 Bus - If using the IEEE488 bus, a Keithley, CEC, or National Instruments IEEE-488.2 interface board must be installed in the PC. Use an IEEE488.2 cable to connect the interface board to the Model 6487 mainframe. RS-232 interface - If using the RS-232 port, use an RS-232C cable to connect the communications port to the Model 6487. When using RS-232 interface, the RS-232 cable must be a straight-through cable, not a null modem cable. Perform the following steps: Download the 6487 A02 Firmware upgrade.exe file from the Keithley Web site, www.keithley.com to a temporary directory on your PC. From Windows, click Start>>Run>>Browse. From the dialog box, navigate to the temporary directory and select the 6487 A02 Firmware upgrade.exe file. Click OK or Open. This will unpack the files required to install the flash upgrade utility. From Windows, click Start>>Run>>Browse. From the dialog box, navigate to the temporary directory and select the Setup.exe file. Click OK or Open. Follow all the prompted instructions to install the upgrade utility. From Windows, click Start>>Programs>>Keithley Instruments>>Model 6487 Flash Upgrade A02, and follow all prompted instructions. Known Issues: When using RS-232 interface, The user may get a timeout error at 99% (after instrument resets): "Timeout Error: Error In Programming Model 6487!". This error could be safely ignored. --------------------------------------------------- End of 6487 A02 Firmware Upgrade.txt